Troubleshooting

Please choose a troubleshooting subject from the list below.
If an answer to your question does not appear, you may contact BioStar Technical Services at 1(800) 637-3717.

Unexpected Patient Results

  1. The results show an unexpected increase in pH.
  2. False positive (+) Glucose result.
  3. False positive (+) Blood result.
  4. False positive (+) Leukocytes result.
  5. False positive (+) Protein result.
  6. False negative (–) Glucose result.
  7. False negative (–) Blood result.
  8. False negative (–) Leukocytes result.
  9. The PocketChem result and my visual interpretation do not agree.

Error Codes

  1. E001 - Optical Movement
  2. E002 - Lamp Intensity High
  3. E003 - Lamp Intensity Low
  4. E004 - Out of Paper
  5. E005 - Printer
  6. E006 - Strip Error

Problems with Interpretation and Quality Control

  1. Erroneous patient result reading strip visually.
  2. Erroneous patient result using PocketChem UA Analyzer.
  3. Check Strip results do not match those on the Check Strip container.
  4. Quality Control result is out of range.

Miscellaneous Equipment Problems

  1. Cannot access Special Parameters menu.
    (Test Strip Holder orientation, Date layout, and Measurement Units)
  2. Test Strip Holder will not fit back into testing position in Analyzer.
  3. Blinking Battery icon.
  4. PocketChem instrument will not turn on.
  5. Keypad or Display screen not responding.
  6. There is no printing on the paper print-outs.


Problem Possible Cause Solution
Results show an unexpected increase in pH. Urine at room temperature for longer than 1 hour causing overgrowth of organisms (if present) with increase in pH and peroxidase production. Refrigerate specimen if not tested within 1 hour.
False Positive (+) Glucose result.

1. Oxidizing substances such as hypochlorite and chlorine may be present in the sample
2. The urine is acidic (pH < 4).

N/A
False Positive (+) Blood result. Urine at room temperature for longer than 1 hour causing overgrowth of organisms (if present) with increase in pH and peroxidase production. Refrigerate specimen if not tested within 1 hour.
False Positive (+) Leukocytes result.

1. There may be formaldehyde in the sample
2. Female patients may occasionally have positive results due to contamination of sample with vaginal discharge.

Recollect specimen.
False Positive (+) Protein result. 1. Urine at room temperature for longer than 1 hour causing overgrowth of organisms (if present) with increase in pH and peroxidase production Refrigerate specimen if not tested within 1 hour.
  2. Specimen may contain disinfectants, antiseptics, detergents, or skin cleanser compunds containing Chlorhexidine. Collect specimen without use of such skin cleansers and disinfectant solutions.
Problem Possible Cause Solution

False negative (–) Glucose result.

1. Large amounts of ascorbic acid (>25 mg/dL) in sample
2. Urine with significantly high Specific Gravity.
N/A
False negative (–) Blood result.

1. Large amounts of ascorbic acid (>25 mg/dL) in sample
2. Capoten (Captopril™) in sample
3. Urine with elevated Specific Gravity or Protein
4. Oxidizing substances such as hypochlorite and chlorine in sample.

N/A
False negative (–) Leukocytes result.

1. Glucose >500 mg/dL
2. Protein >300 mg/dL
3. Urine with low pH (<5)
Tetracycline in high levels in sample
4. Cephalixin (Keflex™), cephalothin (Keflin™), or high concentrations of axalic acids in sample.

N/A
The PocketChem result and my visual interpretation do not agree. There may be variability between visual interpretation of the results and that of the analyzer due to differences between the human eye and optical systems of the analyzer. Refer to the AUTION Sticks 10TA Package Insert for detectable levels of analytes. Each color block represents a range of values. Specimens with analyte concentrations that fall between nominal levels may give results at either level. Results at levels greater than +1 for Glucose, Ketones, Protein, and Urobilinogen will be within 1 level of the true concentration.
Problem Possible Cause Solution

E001 - Optical Movement

1. Optical system does not function correctly Place the instrument on a stable and level surface free of vibration when running.
2. Movement of instrument during measurement Make sure Test Strip Holder and Test Strip are set properly and turn on power. Switch Strip type to 'C' and perform Check Strip test.
E002 - Lamp Intensity High Lamp intensity of LED is too strong Clean the white plate of the Test Strip Holder with a soft cloth moistened with water or alcohol and turn on power. Switch Strip type to 'C' and perform Check Strip test.
E003 - Lamp Intensity Low Lamp intensity of LED is too weak Clean the white plate of the Test Strip Holder with a soft cloth moistened with water or alcohol and turn on power. Switch Strip type to 'C' and perform Check Strip test.
E004 - Out of Paper There is no paper in printer, or it is not correctly installed Lift paper cover, drop a new roll of thermal paper, with paper feeding from the bottom of roll, into the paper holder and feed leading edge into lower slot of black paper feed. If the instrument is plugged in, the leading edge will feed automatically. You can also press the FEED button to advance the paper.
E005 - Printer Printer not functioning correctly 1. Check connections between analyzer and printer. Check ports for dirt, grit, or lint causing a poor connection.
2. Check AC Adapter
3. Press STOP to restore the Ready Screen
E006 - Strip Error The test strip is not inserted correctly Insert the Test Strip correctly by placing it on the Strip Holder until the reagent end of the test strip hits the end of the strip holder. The handle end of the reagent strip will extend about ½" outside of the strip holder. To obtain measurement, a new strip must be analyzed.
Problem Possible Cause Solution

Erroneous patient result reading strip visually.

1. Color changes may occur after 2 minutes Proper read time is critical for optimal results. All analytes should be visually read at 60 seconds, except Leukocytes, which are read at 90 seconds. Color changes that occur after 2 minutes are of no diagnostic value. Read the strips in a well-lit area avoiding direct sunlight.
2. Using deteriorated, discolored, or blackened test strips Always replace the cap of reagent storage bottle tightly after removing a strip for use. Use the stick immediately. All unused sticks must remain in the original bottle. Transfer to any container may cause reagents on the sticks to deteriorate and become unreactive. Do not remove desiccant(s) from bottle.
3. AUTION Sticks 10TA have expired Do not use expired Test Sticks.
Erroneous patient result using PocketChem UA Analyzer. 1. Not using AUTION Sticks 10TA Use only AUTION Sticks 10TA with PocketChem UA analyzer.
2. Instrument problem a. When running routine urinalysis with AUTION Sticks, Test Strip type in analyzer must be set to "10TA."
b. Confirm Performance of AUTION Sticks with external Quality Control measurement.
c. Confirm Performance of PocketChem UA with Check Strip measurement.
Check Strip results do not match those on the Check Strip container. 1. Test Strip Type was not selected and changed to "C" before running Check Strip measurement Follow procedure in User Manual to change Test Strip Type to "C" from "10TA" before running Check Strip measurement.
2. Gray areas on Check Strip are contaminated Do not touch gray areas on Check Strip to prevent contamination.
3. Defective Check Strip Rerun using new Check Strip.
4. Instrument problem Troubleshoot possible instrument problem.
Quality Control result is out of range. 1. Quality Control material problem Check storage, expiration date, and correct control material being used. Repeat measurement with fresh control material.
2. Instrument problem Confirm instrument performance with Check Strip measurement. If Check Strip measurement is within range repeat control measurement. If Check Strip measurement is out of range, repeat with new Check Strip. If still out of range call Biostar Technical Support at 800-637-3717, #2.
3. Not using AUTION Sticks 10TA Use only AUTION Sticks 10TA with PocketChem UA analyzer.
Problem Possible Cause Solution

Cannot access Special Parameters menu (Test Strip Holder orientation, Date layout, and Measurement Units)

STOP key not held down while powering-on analyzer The STOP key must be depressed during power-on AND initialization to open Special Parameter screen. Once Special Parameter screen is open, the blinking digits on right can be changed and saved by pressing ENTER after selection.
Test Strip Holder will not fit back into testing position in Analyzer Guide on bottom of Test Strip Holder not in track Place the guide on bottom of Test Strip Holder into track on bottom of Optical System and slide into place. Make sure the Holder snaps back into place completely.
Blinking Battery icon. Batteries have been depleted Replace batteries with 2 "AA" batteries. Analyzer unit must be removed from printer to access the battery compartment on the back of the analyzer. Remove old batteries and replace with fresh ones. The printer dock is NOT capable of recharging the batteries. Batteries will only power the analyzer unit. When using the analyzer with the printer the AC Adapter must be used for power.
PocketChem instrument will not turn on. 1. AC Adapter problem Check all AC adapter connections. Replace AC Adapter (can be ordered from Biostar).
2. Batteries depleted Replace old batteries with fresh ones, or try docking the Analyzer to the plugged-in Printer.
3. Instrument has been physically damaged Check all connections on the back of instrument for dirt, lint or other grit. Turn power ON again by pressing ENTER. In the event of severe damage, contact BioStar regarding warranty or replacement options.
4. Instrument has been exposed to chemicals or other liquids or electronic noise Do not place instrument where it can be affected by direct sunlight, liquids, electronic noise, or other magnetic material.
Problem Possible Cause Solution
Keypad or Display screen not responding. Possible software problem Hold down the START key for 5 seconds to power-off instrument, then power it on again. If this does not work, remove the Analyzer from the Printer and remove the batteries. Replace batteries and power-on again. Results in memory will NOT be lost.
There is no printing on the paper print-outs. Thermal paper loaded incorrectly Take thermal printer paper out and load correctly. Roll should feed from the bottom with smooth side of paper facing operator when printing.